TRIGger:PULSe Subsystem
TVS600 & TVS600A Command Reference
2–281
TRIGger:PULSe:WIDTh:QUALify
TRIGger:PULSe:WIDTh:QUALify?
Sets or queries the type of time qualification for pulse width triggering. The type
IN triggers when the measured pulse width is within the specified lower and
higher limits. OUT triggers when the pulse width is outside the specified lower
and upper limits.
This command is active only when you set TRIGger:TYPE to PULSe and
TRIGger:PULSe:CLASs to WIDTh.
TRIGger[:A]:PULSe:WIDTh:QUALify <type>
TRIGger[:A]:PULSe:WIDTh:QUALify?
<type> Query response
IN
OUT
IN
OUT
IN
Execution Error –224, “Illegal parameter value”
Attempted to set qualify to an illegal value.
None
Command: TRIG:PULS:WIDT:QUAL OUT
Query: TRIG:PULS:WIDT:QUAL?
Response: OUT
TRIGger:TYPE
TRIGger:PULSe:CLASs
TRIGger:PULSe:WIDTh:LLIMit
TRIGger:PULSe:WIDTh:HLIMit
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands