Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
AVERage Subsystem
TVS600 & TVS600A Command Reference
2–15
AVERage:COUNt
AVERage:COUNt?
Sets or queries the number of acquisition records to average. Averaging reduces
signal noise by approximately 3 dB for each power of 2 increase (2, 4, 8, 16) in
the value of COUNt. For example, a COUNt setting of 8 will result in 3 dB less
noise than a COUNt of 4. The COUNt setting affects all active channels. The
count setting of one is intended for the PEAK DET type of averaging. PEAK
DET uses hardware to envelope the acquired data. The other modes will function
with a count of one, but with no added value.
[SENSe:]AVERage:COUNt <count>
[SENSe:]AVERage:COUNt?
<count> Query response
<NRf>
1 N 4096
MINimum
MAXimum
<NR1>
2, 1
1
4096
1
TVS600A only
2
Execution Error –222, “Data out of range”
Attempted to set count to an illegal value.
None
Command: AVER:COUN 16
Query: AVER:COUN?
Response: 16
AVERage
AVERage:TYPE
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands