FUNCtion and DATA Subsystems
TVS600 & TVS600A Command Reference
2–109
FUNCtion:STATe
FUNCtion:STATe?
Sets or queries the state of a specified sense function. It operates like the
FUNCtion[:ON] and FUNCtion:OFF commands which allow you to enable or
disable a function.
Channel names and function strings may be used interchangeably for parameters.
This is not standard SCPI practice and does not provide horizontal compatibility.
[SENSe:]FUNCtion:STATe <function>, <boolean>
[SENSe:]FUNCtion:STATe? <function>
<function> Query response
“XTIMe:VOLTage[:DC] 1” or CHAN1
“XTIMe:VOLTage[:DC] 2” or CHAN2
“XTIMe:VOLTage[:DC] 3” or CHAN3
“XTIMe:VOLTage[:DC] 4” or CHAN4
Not applicable
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
0
Execution Error –224, “Illegal parameter value”
Attempted to to enable or disable an invalid SENSe function.
Execution Error –241, “Hardware missing”
Attempted to enable or disable “XTIM:VOLT 3” or “XTIM:VOLT 4” when the
instrument is configured with two channels.
This command determines the number of records returned by the query DATA?.
Command: FUNC:STAT XTIM:VOLT 1", ON
Query: FUNC:STAT? XTIM:VOLT 1"
Response: 1
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples