Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
VOLTage Subsystem
2–308
TVS600 & TVS600A Command Reference
<offset> (Value subtracted from signal)
1
Query response
PTPeak 10.1 V – 100 V
±100.0 (in 100 mV steps)
MINimum
MAXimum
–100.0E+0
100.0E+0
1
The default multiplier for the <offset> parameter is V for volts. You can also use MV
for millivolts.
0.0E+0
Execution Error –222, “Data out of range”
Attempted to set offset to an illegal value.
Execution Error –241, “Hardware missing”
Attempted to program VOLTage3 or VOLTage4
on an instrument configured with two channels.
VOLTage:RANGe:PTPeak and VOLTage:RANGe:OFFSet operate as a pair.
Changing the voltage range, changes VOLTage:RANGe[:UPPer] and VOLT-
age:RANGe:LOWer so they describe the new voltage range.
Command: VOLT1:RANG:OFFS 1
Query: VOLT1:RANG:OFFS?
Response: 1.0E+0
VOLTage:RANGe:PTPeak
VOLTage:RANGe:LOWer
VOLTage:RANGe[:UPPer]
Reset Value
Errors and Events
Dependencies
Examples
Related Commands