Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–44
TVS600 & TVS600A Command Reference
Execution Error –222, “Data out of range”
Attempted to set transition width to an illegal value.
None
Command: CALC1:FILT:FREQ:TWID 0.2
Query: CALC1:FILT:FREQ:TWID?
Response: 0.2
CALCulate:FILTer:FREQuency:SREJection
CALCulate:FORMat
CALCulate:FORMat?
Sets or queries whether to process the acquisition record to produce a new format
from complex transform records. The available format types are:
H COMPlex—produce an xy pair for each sample point.
H MLINear—calculate the square root of the sum of the squares of x and y.
H MLOGarithmic—calculate 20*log
10
of the square root of the sum of the
squares of x and y.
H NONE—produces no change in the acquisition record data.
H PHASe—calculates the arctan of y over x.
H POLar—calculates a pair of points r1 and r2, where r1 is the square root of
the sum of the squares of x and y and r2 is the arctan of y over x.
Setting format to NONE effectively disables the function :FORMat.
When the source waveform is not complex and format is set to something other
than NONE, the value for y in x +jy is assumed to be zero.
CALCulate<n>:FORMat <format>
CALCulate<n>:FORMat?
Errors and Events
Dependencies
Examples
Related Commands
Syntax