SYSTem Subsystem
2–196
TVS600 & TVS600A Command Reference
SYSTem:COMMunicate:SERial:PACE
SYSTem:COMMunicate:SERial:PACE?
Sets or queries whether software flow control (XON/XOFF) is enabled for the
front panel RS-232 port. Both transmit and receive values are set by this
command. Refer to the TVS600 & TVS600A Series Waveform Analyzers User
Manual for the factory setting.
When you set :SERial:PACE, the new value is stored in nonvolatile RAM and is
recalled at power up. However, the flow control mode is not saved with stored
settings (*LRN) and it is not reset to a default value by *RST.
You should not use software flow control when transferring binary data.
SYSTem:COMMunicate:SERial:PACE <flow_control>
SYSTem:COMMunicate:SERial:PACE?
<flow_control> Query response
NONE
XON
NONE
XON
Not applicable
Execution Error –141, “Invalid character data”
Attempted to set the software flow control to an illegal value.
None
Command: SYST:COMM:SER:PACE XON
Query: SYST:COMM:SER:PACE?
Response: XON
SYSTem:COMMunicate:SERial:CONTrol:DCD
SYSTem:COMMunicate:SERial:CONTrol:RTS
SYSTem:COMMunicate:SERial:PRESet:RAW
SYSTem:COMMunicate:SERial:PRESet:TERMinal
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands