Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:B Subsystem
2–256
TVS600 & TVS600A Command Reference
TRIGger:B:SLOPe
TRIGger:B:SLOPe?
Sets or queries whether triggering occurs on the positive-going or negative-going
edge of the trigger B signal source. This command has effect only when
TRIGger:B:SOURce is set to INTernal or EXTernal. The other trigger sources
are digital signals.
TRIGger:B:SLOPe <trigger_slope>
TRIGger:B:SLOPe?
<trigger_slope> Query response
POSitive
NEGative
POS
NEG
POS
Execution Error –141, “Invalid character data”
Attempted to set slope to an illegal value.
None
Command: TRIG:B:SLOP NEGATIVE
Query: TRIG:B:SLOP?
Response: NEG
TRIGger:B:LEVel
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands