Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
SWEep Subsystem
2–180
TVS600 & TVS600A Command Reference
SWEep:TINTerval
SWEep:TINTerval?
Sets or queries the time interval between acquired data points. The reciprocal of
the time interval gives the effective sample rate. All channels are acquired with
the same time interval.
The point at which the instrument transitions from real time (RT) acquisition to
extended real time (ERT) acquisition mode is independent of instrument
configuration or settings.
RT < 100 ns ERT
[SENSe:]SWEep:TINTerval <time_interval>
[SENSe:]SWEep:TINTerval?
<time_interval> (seconds)
1
Query response
<NRf>
200E–12 (TVS625A, TVS645A only)
400E–12 (TVS625A, TVS645A only)
1E–9
2E–9
4E–9
10E–9
20E–9
40E–9
100E–9
.
.
.
200E–3
MINimum
MAXimum
<NR3>
200.0E–12
400.0E–12
1.0E–9
2.0E–9
4.0E–9
10.0E–9
20.0E–9
40.0E–9
100.0E–9
.
.
.
200.0E–3
depends on model
200.0E–3
1
The default multiplier for <time_interval> is S for seconds. You can also use the
multipliers MS for milliseconds, US for microseconds, NS for nanoseconds, and PS
for picoseconds.
1.0E–9
Execution Error –222, “Data out of range”
Attempted to set time interval to an illegal value.
Execution Error –241, “Hardware missing”
Attempted to set time interval to 200E–12 or 400E–12 when the instrument is a
TVS621 or TVS641.
Syntax
Parameters
Reset Value
Errors and Events