Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALibration Subsystem
TVS600 & TVS600A Command Reference
2–91
CALibration:RESults:VERBose? (Query Only)
Returns an ASCII string describing the results of the last calibration performed.
The returned string will include the numeric identifier for the first function that
failed. When there are no failures, the query returns a zero value and a descrip-
tion of the last executed function. When a failure occurs, the query returns the
number of the first failed function followed by detailed results information. If the
query returns a value of –1, calibration is in progress.
Calibration functions have numeric identifiers in the range 2000 to 2999.
Because this query is not synchronized with instrument operations so you may
want to use *WAI, *OPC, or *OPC? to wait for the CAL_OPC pending flag to
clear before sending the CAL:RES? query.
CALibration:RESults:VERBose?
Query response
Not applicable <NR1>,<string>
<NR1>:results code
–1 (Calibration in progress)
0 (No failures)
2000 N 2999
<string>:verbose results
(error specific)
Not applicable
None
None
Query: CAL:RES:VERB?
Response: "2001,CAL_1 max=444 min=222 . . ."
CALibration[:ALL]
CALibration:RESults?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands