Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger[:A] Subsystem
TVS600 & TVS600A Command Reference
2–243
Changing the trigger source from INTernal to EXTernal or from one INTernal
source to another INTernal source with a different VOLTage:RANGe setting may
change the TRIGger:LEVel setting to accommodate a new vertical range.
Command: TRIG:SOUR INT1
Query: TRIG:SOUR?
Response: INT1
ARM:SOURce
TRIGger:B:SOURce
TRIGger:TYPE
TRIGger:TYPE?
Sets or queries the type of triggering to use for the next acquisition. The
available types are as follows:
H EDGE specifies the default triggering. A trigger event occurs when a signal
passes through a specified voltage level on the specified rising or falling
slope. The standard SCPI trigger commands control this mode.
H LOGic–A trigger event occurs when the specified combination of input
signals occur. (TVS600A models only)
H PULse generates a trigger event when a specified pulse is detected. It is
controlled by commands in the TRIGger[:A]:PULSe subsystem.
H SHOLdtime–A trigger event occurs when the specified clock-to-data setup
and hold time is violated. (TVS600A models only)
H TRANsition–A trigger event occurs when the input signal fails to transition
between two thresholds in the specified time. (TVS600A models only)
TRIGger[:A]:TYPE <trigger_type>
TRIGger[:A]:TYPE?
<trigger_type> Query response
EDGE
LOGic
PULSe
SHOLdtime
TRANsition
EDGE
LOG
PULS
SHOL
TRAN
Dependencies
Examples
Related Commands
Syntax
Parameters