TRIGger:LOGic Subsystem
TVS600 & TVS600A Command Reference
2–261
TRIGger:TYPE
TRIGger:LOGic:CONDition
TRIGger:LOGic:CONDition?
TVS600A Models Only
Sets or queries the trigger condition for each channel when logic triggering is
selected. The pattern argument refers to the two or four channels of the instru-
ment in a left to right syntax for channel 4(2) through channel 1. A “1” indicates
a High, a “0” indicates a Low, and an “X” indicates a Don’t Care is selected.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is set to LOGic.
The condition of channel 4 is ignored when TRIGger:LOGic:CLASs is set to
STATe.
The condition of Bits 3 and 4 are ignored when the instrument is configured with
two channels.
TRIGger[:A]:LOGic:CONDition <pattern>
TRIGger[:A]:LOGic:CONDition?
<paggern> Query response
LCXXX0
LCXXX1
....
LCM1111
LCXXX0
LCXXX1
LCM1111
None
LC0000
Command Error –141, “Invalid character data”
Attempted to set either condition parameter to an illegal value.
None
Command: TRIG:LOG:COND LC1100
Related Commands
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples