Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–40
TVS600 & TVS600A Command Reference
<start> (Beginning frequency of range)
1
Query response
<NRf> <NR3>
1
The default multiplier for the parameter <start> is HZ for hertz. You can also use the
multipliers KHZ for kilohertz, MHZ for megahertz, and GHZ for gigahertz.
200.0E+6
Execution Error –222, “Data out of range”
Attempting to set start to an illegal value will generate execution error.
You can set the frequency range for the frequency filters :NOTCh and :BPASs
with either the command pair :CENTer and :SPAN or the command pair :STARt
and :STOP. Because both command pairs control the same parameters, changing
either command pair will affect the other.
Command: CALC1:FILT:FREQ:STAR 9.99E6
Query: CALC1:FILT:FREQ:STAR?
Response: 9.99E+6
CALCulate:FILTer:FREQuency:STOP
CALCulate:FILTer:FREQuency:CENTer
CALCulate:FILTer:FREQuency:SPAN
Reset Value
Errors and Events
Dependencies
Examples
Related Commands