VOLTage Subsystem
TVS600 & TVS600A Command Reference
2–309
VOLTage:RANGe:PTPeak
VOLTage:RANGe:PTPeak?
Sets or queries the peak-to-peak (full-scale) voltage range of the specified input
amplifier. The range of permitted values and the smallest increment for offset
depend on the setting of the peak-to-peak range (VOLTage:RANGe:PTPeak).
For specific values, see the listing of <range> values under Parameters.
The values permitted for <range> also depend on the attached probe. An attached
probe modifies the full-scale range by the attenuation factor of the probe. For
example, a 10X probe changes the 10 mV range to the 100 mV range.
[SENSe:]VOLTage<n>[:DC]:RANGe:PTPeak <range>
[SENSe:]VOLTage<n>[:DC]:RANGe:PTPeak?
<n> (Channel number)
1
Query response
1
2
3
4
Not applicable
1
The input channel number <n> is used by the commands INPut<n> and VOLTage<n>.
If you omit <n>, the default is channel 1.
<range> (Range for vertical amplifier)
1
Query response
<NRf> <NR3>
10 mV – 20 mV (100 mV steps)
20 mV – 50 mV (200 mV steps)
50 mV – 100 mV (500 mV steps)
100 mV – 200 mV (1 mV steps)
200 mV – 500 mV (2 mV steps)
500 mV – 1 V (5 mV steps)
1 V – 2 V (10 mV steps)
2 V – 5 V (20 mV steps)
5 V – 10 V (50 mV steps)
10 V – 20 V (100 mV steps)
20 V – 50 V (200 mV steps)
50 V – 100 V (500 mV steps)
MINimum
MAXimum
10.0E–3
100.0E+0
1
The default multiplier for the <offset> parameter is V for volts. You can also use MV
for millivolts.
1.0E+0
Syntax
Parameters
Reset Value