TRIGger:LOGic Subsystem
TVS600 & TVS600A Command Reference
2–263
TRIGger:LOGic:FUNCtion
TRIGger:LOGic:FUNCtion?
TVS600A Models Only
Sets or queries the logical combination of the input channels for the main logic
trigger. When TRIGger:LOGic:CLASs is set to PATTern, this command applies
to channel 1, 2, 3 and 4. When TRIGger:LOGic:CLASs is set to STATe, only
channels 1, 2 and 3 are logically combined.
:A is a pre–defined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is set to LOGic.
TRIGger[:A]:LOGic:FUNCtion <logic_func>
TRIGger[:A]:LOGic:FUNCtion?
<logic_fun> Query response
AND
NAND
NOR
OR
AND
NAND
NOR
OR
None
AND
Command Error –141, “Invalid character data”
Attempted to set FUNCtion to an illegal value.
None
Command: TRIG:LOG:FUNC NOR
Query: TRIG:LOG:FUNC?
Response: NOR
TRIGger:TYPE
TRIGger:LOGic:CLASs
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands