Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:TRANsition Subsystem
TVS600 & TVS600A Command Reference
2–301
TRIGger:TRANsition:THReshold:LOW
TRIGger:TRANsition:THReshold:LOW?
TVS600A Models Only
Sets or queries the low threshold for transition triggering.
:A is a predefined alias for :SEQuence[1].
The minimum and maximum values for threshold are VOLT:RANG:OFFS +/–
VOLT:RANG:PTP (in VOLT:RANG:PTP * 0.002 steps).
This command only has effect when TRIGger:TYPE is TRANsition.
TRIGger[:A]:TRANsition:THReshold:LOW <level>
TRIGger[:A]:TRANsition:THReshold:LOW?
<level> Query response
<NRf>
MINimum
MAXimum
<NR3>
<See comments>
<See comments>
V (default), MV, UV
0.0E+0
Execution Error –222, “Data out of range”
Attempted to set threshold to an illegal value.
None
Command: TRIG:TRAN:THR:LOW -1.0
Query: TRIG:TRAN:THR:LOW?
Response: -1.0E+0