CALCulate Subsystem
2–20
TVS600 & TVS600A Command Reference
SENSe
CALCulate1..4
TRACe
CHAN1..4
”XTIM:VOLT n”
”CALCn”
:PATH
CALC1..4
TRACe
AATS
:PATH:EXPRession
:FEED1
:FEED2
Figure 2–8: CALCulate subsystem functional model
CALCulate:AAMList
CALCulate:AAMList?
Sets or queries the list of measurements to perform on the selected auto-advance
acquisition records. You can specify up to 50 measurements by separating them
with commas. Before you can make auto-advance measurements, you must
enable the auto-advance measurement system with the command CALCu-
late:AAMList:STATe.
Specify the measurement methods and reference values with the
CALC:WMParameter commands.
The records on which measurements are made are specified by the
AADV:REC:STARt and AADV:REC:COUNt commands.
CALCulate<n>:AAMList <list>
CALCulate<n>:AAMList?
<n>
1
Query response
1
2
3
4
NA
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
Syntax
Parameters