Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–76
TVS600 & TVS600A Command Reference
CALCulate:WMParameter:HREFerence
CALCulate:WMParameter:LREFerence
CALCulate:WMParameter:LREFerence?
Sets or queries the low reference (proximal) level in vertical units for time and
amplitude measurements. This setting is used only when
CALC:WMP:RMEThod is set to ABSolute.
CALCulate<n>:WMParameter:LREFerence[:ABSolute] <lrefā>
CALCulate<n>:WMParameter:LREFerence[:ABSolute]?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<lref> (Set absolute value of LREF) Query response
<NRf> <NR2>
0.0E+0
Execution Error –222, “Data out of range”
Attempted to set the low reference to an illegal value.
None
Command: CALC1:WMP:LREF -8
Query: CALC1:WMP:LREF?
Response: -8.0E+0
CALCulate:WMParameter:LREFerence:RELative
CALCulate:WMParameter:RMEThod
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands