CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–63
CALCulate:WMParameter:GATE
CALCulate:WMParameter:GATE?
TVS600A Models Only
Sets or queries the state of measurement gating. Settings established with other
GATE parameters are ignored unless CALC:WMP:GATE is set to ON.
CALCulate[1]:WMParameter:GATE <boolean>
CALCulate[1]:WMParameter:GATE?
.
.
CALCulate4:WMParameter:GATE <boolean>
CALCulate4:WMParameter:GATE?
<boolean> Query response
<NRf>
N ≠ 0
0
ON
OFF
<NR1>
1
0
1
0
None
0
None
None
Command:
CALC1:WMP:GATE ON
Query: CALC1:WMP:GATE?
Response: 1
CALCulate:WMParameter:GATE:MODe
CALCulate:WMParameter:GATE:STARt
CALCulate:WMParameter:GATE:STOP
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands