Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
AVERage Subsystem
2–16
TVS600 & TVS600A Command Reference
AVERage:TYPE
AVERage:TYPE?
Sets or queries the mode of acquisition to perform:
H SCALar (average), the default, averages each new sample point with the
corresponding point in the previous acquisition record.
H ENVelope generates a waveform record of interleaved MAX/MIN pairs of
sample points (the number of pairs is SWEep:POINts divided by two).
H PEAKdetect (TVS600A only) generates an array of MAX/MIN pairs, but
uses hardware peak detect to capture glitches between sampling intervals.
The AVERage:TYPE setting affects all channels.
Peak Detect hardware is not functional at a sample rate faster than 100 MS/s.
When peak detect is enabled on rates faster than 100 MS/s, the instrument will
function as if in ENVelope mode and will not issue any warning or error to that
effect.
[SENSe:]AVERage:TYPE <type>
[SENSe:]AVERage:TYPE?
<type> Query response
ENVelope
SCALar
PEAKdetect
1
ENV
SCAL
PEAK
1
1
TVS600A only
Syntax
Parameters