CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–79
CALCulate:WMParameter:MREFerence:RELative
CALCulate:WMParameter:RMEThod
CALCulate:WMParameter:MREFerence:HYSTeresis
CALCulate:WMParameter:MREFerence:HYSTeresis?
Sets or queries the middle reference (mesial) hysteresis used for time and
amplitude measurements. The hysteresis can be expressed as a ratio or percent of
the current value for AMPLitude. The signal must transition beyond the vertical
range defined by :MREFerence:HYSTeresis before another crossing of
:MREFerence can be accepted for measurements such as :PERiod.
This setting is used only when CALC:WMP:RMEThod is set to RELative.
CALCulate<n>:WMParameter:MREFerence:HYSTeresis <hyst>
CALCulate<n>:WMParameter:MREFerence:HYSTeresis?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<hyst> (Set relative value of HYST) Query response
<NRf>
0.0 v N v 0.5
1
MINimum
MAXimum
<NR2>
0.0
0.5
1
The units must be specified as PCT when setting MREFerence:HYSTeresis as a
percent of the value AMPL.
0.05
Execution Error –222, “Data out of range”
Attempted to set the middle reference hysteresis to an illegal value.
None
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies