Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
AADVance Subsystem
2–6
TVS600 & TVS600A Command Reference
AADVance:RECord:COUNt
AADVance:RECord:COUNt?
Sets or queries the number of auto-advance acquisition records to transfer in
response to the commands DATA?, TRACe?, TRACe:COPY?, or
TRACe:LIST?. The maximum :COUNt value depends on the number of
acquired records and the value of AADVance:RECord:STARt. A setting of
zero (0) selects all acquisition records, beginning with :STARt, for transfer
regardless of the current acquisition system settings. MAXimum selects all
acquisition records for transfer based on the current acquisition settings. If you
change a setting, such as the number of active channels, the MAXimum setting
will not be adjusted, but a setting of zero will automatically adjust for the new
acquisition settings.
[SENSe:]AADVance:RECord:COUNt <count>
[SENSe:]AADVance:RECord:COUNt?
<count> Query response
<NRf>
0 (Transfer all records)
1 to MAX (Transfer number of
records)
MINimum
MAXimum
<NR1>
1
<depends on configuration>
1
Execution Error –222, “Data out of range”
Attempted to set count to an illegal value.
None
Command: AADV:REC:COUN 100
Query: AADV:REC:COUN?
Response: 100
AADVance
AADVance:COUNt
AADVance:RECord:STARt
CALC:AAML
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands