Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger:TRANsition Subsystem
2–294
TVS600 & TVS600A Command Reference
TRIGger:TRANsition:CLASs
TRIGger:TRANsition:CLASs?
TVS600A Models Only
Sets or queries the class of transition triggering selected. RUNT: the instrument
triggers when a pulse crosses one of two thresholds but does not cross the second
threshold before recrossing the first. SLEWrate: the instrument triggers when an
edge crosses two thresholds. All classes may be time qualified.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is TRANsition.
TRIGger[:A]:TRANsition:CLASs <transition_class>
TRIGger[:A]:TRANsition:CLASs?
<transition_class> Query response
RUNT
SLEWrate
RUNT
SLEW
None
RUNT
Command Error –141, “Invalid character data”
Attempted to set class to an illegal value.
None
Command: TRIG:TRAN:CLAS SLEWRATE
Query: TRIG:TRAN:CLAS?
Response: SLEW
TRIGger:TYPE
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands