FUNCtion and DATA Subsystems
2–112
TVS600 & TVS600A Command Reference
DATA:PREamble? (Query Only)
Returns the data preamble for the specified function or for all enabled sense
functions. If auto-advance acquisition is on, then one preamble for each enabled
channel is returned.
When you specify a sense function, DATA:PREamble? returns only the preamble
of the specified function. When no sense function is specified,
DATA:PREamble? returns the preambles of all enabled sense functions. The
preambles are returned in ascending channel order.
DATA:PREamble? does not return data until pending acquisitions or calculations
are complete. You can use the commands *WAI, *OPC, or *OPC? to synchro-
nize the response in a different manner.
Channel names and function strings may be used interchangeably for parameters.
This is not standard SCPI practice and does not provide horizontal compatibility.
[SENSe:]DATA:PREamble? [<function>]
<function> Query response
“XTIMe:VOLTage[:DC] 1” or CHAN1
“XTIMe:VOLTage[:DC] 2” or CHAN2
“XTIMe:VOLTage[:DC] 3” or CHAN3
“XTIMe:VOLTage[:DC] 4” or CHAN4
Not applicable
<preamble> Query response
Not applicable DIF expression
Not applicable
Execution Error –224, “Illegal parameter value”
Attempted to query data from an invalid sense function.
Execution Error –230, “Data corrupt or stale.”
Attempted to query data that is invalid, incomplete or stale.
Execution Error –241, “Hardware missing”
Attempted to query data from “XTIM:VOLT 3” or “XTIM:VOLT 4” when the
instrument is configured with two channels.
Syntax
Parameters
Reset Value
Errors and Events