Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
SYSTem Subsystem
TVS600 & TVS600A Command Reference
2–203
SYSTem:ERRor:CODE:ALL? (Query Only)
Returns the list of all event codes stored in the waveform analyzer Status Queue.
The returned error codes do not include their associated event messages. Event
codes are separated by commas. Use SYSTem:ERRor:CODE? to return one error
code at a time. Use SYSTem:ERRor:ALL? to return all error codes and event
messages.
Refer to the TVS600 & TVS600A Series Waveform Analyzers User Manual for a
detailed description of the Status Queue and a complete list of system events.
SYSTem:ERRor:CODE:ALL?
<code> Query response
Not applicable <NR1>
Not applicable
0, “No error”
Attempted to read an event when the queue is empty.
Device Specific Error –350, “Queue Overflow”
The Status Queue has overflowed or run out of memory.
None
Query: SYST:ERR:CODE:ALL?
Response: -221,-224
SYSTem:ERRor?
SYSTem:ERRor:CODE
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands