TRIGger:PULSe Subsystem
2–272
TVS600 & TVS600A Command Reference
TRIGger:PULSe:GLITch:QUALify
TRIGger:PULSe:GLITch:QUALify?
Sets or queries the type of time qualification for pulse glitch triggering. LT
specifies that the trigger event occurs when the glitch width is less than the
specified time. GT specifies that the trigger event occurs when the glitch width is
greater than the specified time.
Before this command will be effective, you must set TRIGger:TYPE to PULSe
and TRIGger:PULSe:CLASs to GLITch. Refer to the discussion of TRIG-
ger:PULSe:CLASs on page 2–270.
TRIGger[:A]:PULSe:GLITch:QUALify <type>
TRIGger[:A]:PULSe:GLITch:QUALify?
<type> Query response
GT
LT
GT
LT
LT
Execution Error –224, “Illegal parameter value”
Attempted to set qualify to an illegal value.
None
Command: TRIG:PULS:GLIT:QUAL GT
Query: TRIG:PULS:GLIT:QUAL?
Response: GT
TRIGger:TYPE
TRIGger:PULSe:CLASs
TRIGger:PULSe:GLITch:WIDTh
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands