TRIGger[:A] Subsystem
TVS600 & TVS600A Command Reference
2–235
SWEep:OFFSet:TIME
TRIGger:B:DELay
TRIGger:B:ECOunt
TRIGger:FILTer[:LPASs]
TRIGger:FILTer[:LPASs]?
Sets or queries the state of the 50 kHz low pass trigger filter. Components of the
trigger signal above 50 kHz are attenuated when the LPASs filter is on. It may
only be used when DC coupled and with TRIGger:SOURce set to INTernal.
The INPut:FILTer is separate from the trigger filter and is located before the
trigger pickoff in the signal path.
TRIGger[:A]:FILTer[:LPASs][:STATe] <boolean>
TRIGger[:A]:FILTer[:LPASs][:STATe]?
<boolean> Query response
<NRf>
N ≠ 0 or ON
0 or OFF
<NF1>
1
0
0
None
Enabling the low pass filter will set TRIGger:COUPling to DC and set TRIG-
ger:FILTer:NREJect to OFF. When TRIGger:A and TRIGger:B share the same
SOURce, changing FILTer for one will change FILTer for the other.
Command: TRIG:FILT ON
Query: TRIG:FILT?
Response: 1
TRIGger:COUPling
TRIGger:FILTer:HPASs
INPut:FILTer
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands