Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
TRIGger[:A] Subsystem
2–236
TVS600 & TVS600A Command Reference
TRIGger:FILTer:HPASs
TRIGger:FILTer:HPASs?
Sets or queries the state of the 50 kHz high pass trigger filter. The HPASs filter
attenuates components of the trigger signal below 50 kHz. HPASs can only be
used when AC coupled and with TRIGger:SOURce set to INTernal.
The INPut:FILTer is separate from the trigger filter and is located before the
trigger pickoff in the signal path.
TRIGger[:A]:FILTer:HPASs[:STATe] <boolean>
TRIGger[:A]:FILTer:HPASs[:STATe]?
<boolean> Query response
<NRf>
N 0 or ON
0 or OFF
<NR1>
1
0
0
None
Enabling the high pass filter will set TRIGger:COUPling to AC and set
TRIGger:FILTer:NREJect to OFF. When TRIGger:A and TRIGger:B share the
same SOURce, changing FILTer for one will change FILTer for the other.
Command: TRIG:FILT:HPAS ON
Query: TRIG:FILT:HPAS?
Response: 1
TRIGger:COUPling
TRIGger:FILTer[:LPASs]
INPut:FILTer
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands