Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALibration Subsystem
TVS600 & TVS600A Command Reference
2–87
CALibration:PROBe
CALibration:PROBe?
TVS600A Models Only
This command calibrates the gain of the attached active voltage probes, but
returns no results code. The query returns a value of zero if the probe is
calibrated.
This command only works with active voltage probes.
Upon successful completion of the probe cal, the calibration constants are stored
nonvolatile along with the probe identification and serial number. These cal
constants are used as long as the identification and serial numbers match the
attached probe.
1. A value of one (1) returned by the query indicates an initialized (uncali-
brated) probe.
2. A value of zero (0) indicates a calibrated probe.
3. A value of minus one (–1) indicates a probe calibration failure.
4. A value of minus two (–2) indicates a probe calibration failed because the
probe was not connected to the calibration source.
5. A value of minus three (–3) indicates a calibration failed because an
unsupported probe or no probe was installed.
The current state of the instrument is saved before the cal and restored after.
This command/query sets the CAL_OPC pending flag. *WAI, *OPC, or *OPC?
may be used to synchronize the command.
CALibration:PROBe[1]
CALibration:PROBe[1]?
CALibration:PROBe2
CALibration:PROBe2?
CALibration:PROBe3
CALibration:PROBe3?
CALibration:PROBe4
CALibration:PROBe4?
Syntax