Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
CALCulate Subsystem
2–38
TVS600 & TVS600A Command Reference
Query: CALC1:FILT:FREQ:SPAN?
Response: 10.0E+3
CALCulate:FILTer:FREQuency:CENTer
CALCulate:FILTer:FREQuency:SREJection
CALCulate:FILTer:FREQuency:SREJection?
Sets or queries the level of rejection or attenuation for frequency components in
the defined stop band for the:FREQuency filters. :SREJection sets the level of
attenuation in dB for the filters :NOTCh, :BPASs, :LPASs and :HPASs. You can
use :FREQuency:TWIDth to set the slope or roll off of the filter.
If you enter too large a value for :SREJection, the function will generate an
execution error and terminate the filter process. If an execution error occurs,
increase the record length or increase the value of :TWIDth. For an overview
discussion of the waveform analyzer digital filter, refer to the TVS600 &
TVS600A Series Waveform Analyzers User Manual.
CALCulate<n>:FILTer[:GATE]:FREQuency:SREJection <level>
CALCulate<n>:FILTer[:GATE]:FREQuency:SREJection?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<level> (Frequency rejection in dB) Query response
<NRf>
15 N 100
MINimum
MAXimum
<NR2>
15.0
100.0
60.0
Related Commands
Syntax
Parameters
Reset Value