Tektronix TVS600 & TVS600A Series Computer Accessories User Manual


 
SYSTem Subsystem
TVS600 & TVS600A Command Reference
2–195
H ^H (backspace),
H ^U (line-delete)
H ^D (end-of-file).
When OFF, characters are immediately passed to the parser, line editing is not
possible, and you must enter a newline (^J) character to terminate a command
line. When transferring binary data, always set the line buffer to OFF. When
buffering is on, return characters (^M) are converted to newlines (^J) and
outgoing newlines (^J) are converted to return-newlines (^M^J). The conversion
of termination characters causes significant problems when transferring binary
data.
When you set :SERial:LBUFfer, the new value is stored in nonvolatile RAM and
is recalled at power up. However, the buffer mode is not saved with stored
settings (*LRN) and it is not reset to a default value by *RST. Refer to the
TVS600 & TVS600A Series Waveform Analyzers User Manual for the factory
setting.
SYSTem:COMMunicate:SERial:LBUFfer <boolean>
SYSTem:COMMunicate:SERial:LBUFfer?
<boolean> Query response
<NRf>
1 or ON
0 or OFF
<NR1>
1
0
Not applicable
None
None
Command: SYST:COMM:SER:LBUF OFF
Query: SYST:COMM:SER:LBUF?
Response: 0
SYSTem:COMMunicate:SERial:PRESet:RAW
SYSTem:COMMunicate:SERial:PRESet:TERMinal
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands