STATus Subsystem
2–166
TVS600 & TVS600A Command Reference
STATus:QUEStionable?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:PTRansition
STATus:PRESet
STATus:QUEStionable:PTRansition
STATus:QUEStionable:PTRansition?
Sets or queries the contents of the Questionable Positive Transition Register
(QPTR). When you set an event bit true (1) in the QPTR, the QPTR reports an
event in the Questionable Status Register when the Questionable Status
Condition Register event line changes from false to true (0 to 1)
. The query
returns the current setting as a decimal number whose bits correspond to the
Questionable Status Register bits. Refer to Table 2–3 on page 2–162 for a
definition of the events associated with the Questionable Status Register bits.
The command STATus:QUEStionable:NTRansition provides similar control for
negative event transitions. For an overview of the status and event reporting
system, refer to the TVS600 & TVS600A Series Waveform Analyzers User
Manual.
STATus:PRESet sets all register bits to one (1) which enables reporting positive
event transitions. Setting unused enable bits does not generate an error, they are
ignored.
STATus:QUEStionable:PTRansition <mask>
STATus:QUEStionable:PTRansition?
<mask> Query response
<NRf> | <Non-decimal numeric>
0 ≤ N ≤ #HFFFF
<NR1>
Not applicable
Execution Error –222, “Data out of range”
Attempted to set the transition register to an illegal value.
None
Command: STAT:QUES:PTR #H0100
With this command, (decimal 256), the Questionable Positive
Transition Register sets the Calibration event bit when calibration
Related Commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples